Typical Installation Sc4A-E-Sa (Sb, Sc) With Standard Options / Spare Parts - YOKOGAWA SC4A Instruction Manual

Specific conductivity sensors
Hide thumbs Also See for SC4A:
Table of Contents

Advertisement

11

2.2.2. Typical installation SC4A-E-SA (SB, SC) with standard options / spare parts

The typical installation of the SC4A-E-SA sensor is done by using a standardized 25 mm port.
Stainless Steel welding sockets are available as straight version (spare part /SA1) or 15° angled
version (spare part /SA2). Both versions are delivered with mounting nut.
In figure 5 an example is shown how to install the sensor using spare part /SA1.
Sensors SC4A-E-SB and SC4A-E-SC are installed by tri-clamp method.
For the SB version two Stainless Steel tri-clamps are available, in a 1" size
(spare part /SB1) or in a 1½" size (spare part /SB2).
The SC version has just one Stainless Steel tri-clamp in a 2" size (spare part /SC1).
In figure 6 an example is shown how to install the sensor using one of these tri-clamps.
Ordering information for spare parts can be found in Section 2.6.
/SA1
Fig. 5 Mounted SC4A-E-SA sensor
Fig. 6 Mounted SC4A-E-SB (SC) sensor
with spare part /SA1 (K1542FA)
with tri-clamp