Installation Instructions; Post-Installation Check - Endress+Hauser StamoLys CA 70 Operating Instructions Manual

Analyzer for photometric measurement
Table of Contents

Advertisement

Installation
10
3.3

Installation instructions

To install the analyzer at the intended location, proceed as follows:
1.
Place the analyzer in position and secure it. For the installation dimensions, please
!
see chapter 3.2.
Note!
A wall bracket is available as an option for mounting the analyzer ( → Chap. 8
"Accessories").
2.
Lay the drain pipe for the reaction products. Where possible, use fixed pipes (PVC
or PE, internal diameter ¾ " with 3% incline).
!
3.
Connect the sample transfer.
Note!
The sample can be obtained as follows:
– Directly or after a reversible flow filter or a cross current filter by means of a small
pump (rating approx. 300 ml/min), suitable for clear media, e.g. in the discharge
channel of a sewage treatment plant
– From a sedimentation tank or after microfiltration; this is practical for media
containing flocculants,
e.g. in an activated sludge basin
– Sample conditioning using ultrafiltration for heavily soiled media,
e.g. from the primary settling tank.
For questions regarding sample conditioning and its automation, please contact
Endress+Hauser Service or the Endress+Hauser Sales Centre responsible for your
region.
4.
Connect the tubes from canisters containing reagents, standard and cleaning
agents to the following nozzles:
Canister
Sample
Reagent 1
Reagent 2
Reagent 3
Standard
Cleaning agent
5.
Tighten the pump hoses onto the sample pump and onto the reagent pump.
Set the pressure at the hose bed throttle valve ( → Fig. 9, P. 35, item D) such that
6.
sample and reagent are fed in without bubbles.
This completes the installation process. Now carry out the installation check and
proceed according to chapters 4 ("Wiring"), 5 ("Operation") and 6 ("Commissioning").
3.4

Post-installation check

• After mounting, check that all connections are secure and leak-tight.
• Ensure that the hoses can not be removed without effort.
• Check all hoses for damage.