Wiring - YOKOGAWA DO72 Instruction Manual

Optical dissolved oxygen sensor
Hide thumbs Also See for DO72:
Table of Contents

Advertisement

5. WIRING

DO72 Sensor cable connection to converter is delivered with wire pin end. For infor-
mation on DO72 sensor cable connection to FLXA402, refer to the user's manual for
FLXA402 4-Wire Converter (IM12A01F01-02EN).
Fig. 9. Connection diagram – WU10-V-D (left) and DO72 sensor wiring details (from
FLXA402 manual - right)
Fig. 10. D6 module terminal assignment for DO72 dissolved oxygen sensor
Table 3: Pin terminal wiring numbering
Wire Pin no.
VP pin
11
G
12
H
13
D
10
C
14
S
Remark: Cables which are not used should be connected to any terminal of some-
where between 15 to 18.
Wire color
Yellow
Green
White (Coax2 shield)
White (Coax2 core)
Black
Description
RS485 Data (A)
RS485 Data (B)
GND
7 – 30 VDC
cable shield / probe body
IM12J06J02-00EN-P
15