COMPACT PIEZO P2K
WARNING: On completion of the
cleaning cycle in the thermal-disinfector,
the scaler handpiece remains at the wash
temperature for a long time. Use appropriate
precautions when extracting the scaler
handpiece from the thermal-disinfector to
prevent injury to the operator.
Position the accessories in a metal basket.
Connect the relevant adapter (supplied
as an optional) to the scaler handpiece
connector and then to the water jet
cleaning connections of the thermal-
disinfector.
Repeat the same operation for the inserts
by connecting them to the appropriate
adapters supplied as optionals.
Sequence and parameters applicable to
the cycle:
• 1 min, Rinse with cold water;
• 5 min, Wash with alkaline detergent
at 55°C ±2°C;
• 1 min, Neutralisation with suitable
solution (1/3 cold water, 2/3 hot
water);
• 1 min, Rinse with water (1/3 cold
water, 2/3 hot water);
• 5 min, Thermal disinfection at 93°C
with demineralised water.
Thermal
disinfection
tested experimentally. In compliance
with ISO 15883-1, Table B.1 [4] thermal
disinfection at a temperature of 90°C for
5 min determines a value of A0 3000.
26
1
2
has
not
been
CAUTION: Owing to its conformation, the
scaler handpiece can rotate. When not in use,
the scaler handpiece must always be placed
on its support.