mectron COMPACT PIEZO P2K Use And Maintenance Manual page 28

Hide thumbs Also See for COMPACT PIEZO P2K:
Table of Contents

Advertisement

COMPACT PIEZO P2K
Leave to soak for 10 minutes at 40°C
±2°C. This procedure reduces the amount
of blood, protein and mucous present on
the insert;
During immersion in the enzymatic
solution, delicately brush the surface of
the insert using the brush with soft nylon
bristles to eliminate all traces of visible
dirt. Thoroughly clean the difficult areas
such as the cutting edges;
Delicately brush the surface of the insert
under running water using the brush with
soft nylon bristles;
Using a syringe, inject the enzymatic
detergent solution 3 times inside the
cavity of the insert to effectively remove
the residues from the internal surface;
22
6
7
8
9