Remove the torque wrench from the
enzymatic detergent solution.
Thoroughly rinse and brush all the
surfaces of the torque wrench (see
previous point) under running water for
at least 10 minutes;
Place the torque wrench in the ultrasonic
tank
immersed
in
detergent solution at 24 °C ±2 °C and run
a cycle for at least 20 minutes;
Remove the torque wrench from the
ultrasonic tank and rinse it under running
water so as to eliminate all detergent
residue.
Brush the internal and external surfaces
of the wrench with a clean brush with soft
nylon bristles, under running water.
6.1.2
Automatic Cleaning
Automatic cleaning can be carried out as an alternative to the manual cleaning described in
Chapter 6.1.1 on page 19
6.1.2.1
Required Materials
• Alkaline detergent: neodisher® FA (0.2
% v/v);
• Neutralising liquid: neodisher® Z (0.1
% v/v);
• Water;
• Metal basket;
• Adapters;
• Thermal-disinfector.
CLEANING AND STERILISATION
17
18
the
enzymatic
19
.
NOTE: Make sure that the accessories are
appropriately blocked in the basket and
cannot move during washing. Any blows
could damage them. Position the instruments
in a way that the water can flow through all
the surfaces, even internal.
WARNING: Avoid overloading the
thermal-disinfector as this could compromise
the effectiveness of cleaning.