Note
This procedure applies to 20 and 28
liter units only.
Note
Do not disconnect internal wiring or
sensors from control panel.
2.
Install the cover by inserting the hinge into the lip at
the rear of the bath. Angle the cover up while push-
ing down to lock the hinge in place under the stain-
less steel tank lip. Remove the cover by lifting the
lid and pushing the hinge towards the back to
release it from under the tank lip. Covers on the 2
liter shallow, 5 liter, and 10 liter model baths can
also be installed and removed by sliding the hinge
into place under the rim and tank lips from either
side.
3.
Connect the line cord to a suitable grounded electri-
cal outlet.

Repositioning the Control Panel

With 20 and 28 liter units, the control panel can be positioned
on the side or the front of the unit in order to achieve the best
use of available bench space. To change the position of the
control panel, follow these steps:
1.
Turn off the unit and disconnect from power source.
2.
Allow to cool.
3.
Remove thermometer and clip. Remove the cover.
4.
Assuming that the water bath is empty, turn it over
on a protected surface.
5.
Remove two screws retaining the control panel and
two screws retaining the blank panel.
6.
Remove the blank panel.
7.
Carefully move the control panel to the new location
and reinstall.
8.
Install the blank panel at the other location.
9.
Return unit to location of use.
10. Reconnect power, turn on and test unit.