Pelstar Health o meter Professional 499KG User Instructions

Pelstar Health o meter Professional 499KG User Instructions

Waist-high digital beam scale
Table of Contents

Advertisement

Available languages

Available languages

Quick Links

499KL / 499KG
Waist-High Digital Beam Scale
Equipped with
®
Everlock
E112697
User Instructions
Para ver las instrucciones de uso en español visita www.homscales.com
Pour les instructions de l'utilisateur dans la visite française www.homscales.com
®
PELSTAR
, LLC
9500 West 55th St. McCook, IL 60525-7110 USA
www.homscales.com
®
Health o meter
Professional 499KL
®
Health o meter
Professional 499KG
© Pelstar, LLC 2020
P/N UM499KL 499KG Rev 20201201_EngOnly

Advertisement

Chapters

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Health o meter Professional 499KG and is the answer not in the manual?

Questions and answers

Summary of Contents for Pelstar Health o meter Professional 499KG

  • Page 1 Pour les instructions de l'utilisateur dans la visite française www.homscales.com ® PELSTAR , LLC 9500 West 55th St. McCook, IL 60525-7110 USA www.homscales.com ® Health o meter Professional 499KL ® Health o meter Professional 499KG © Pelstar, LLC 2020 P/N UM499KL 499KG Rev 20201201_EngOnly...
  • Page 2: Table Of Contents

    ® Health o meter Professional 499KL ® Health o meter Professional 499KG ® Thank you for purchasing this Health o meter Professional product. Please read this manual carefully, and keep it for easy reference or training. TABLE OF CONTENTS Cautions and Warnings ..................... 3 Specifications / Definitions of Symbols ..............
  • Page 3: Cautions And Warnings

    Consult Customer Service or a field service technician for help. In no event whatsoever shall Pelstar, LLC be liable for damages or injuries arising from or connected with the assembly, use, or misuse of its products. P/N UM499KL 499KG Rev 20201201_EngOnly...
  • Page 4: Specifications / Definitions Of Symbols

    SPECIFICATIONS General The model 499KL / 499KG waist high digital beam scale uses highly sophisticated microprocessor technology. This precision scale is designed to provide accurate, reliable and repeatable weight measurements. In addition, each scale is designed to provide the user with features that make the weighing process simple, fast, and convenient.
  • Page 5: Certifications / Connectivity / Disposal

    CERTIFICATIONS / CONNECTIVITY / DISPOSAL Certification Descriptions Pelstar, LLC has been officially certified as the manufacturer of Health o meter® Professional medical devices. The Pelstar, LLC quality assurance system covers the development, production, sales and service of medical scales and measuring systems.
  • Page 6: Assembly Instructions

    ASSEMBLY INSTRUCTIONS Before Assembly Each 499KL / 499KG waist high digital beam scale is shipped disassembled in one carton. Carefully inspect the carton for shipping damage before unpacking. If damage is found, contact ® your shipper or a Health o meter Professional representative immediately at 1-800-815-6615.
  • Page 7 ASSEMBLY INSTRUCTIONS (CONTINUED) 1. Carefully remove the scale platform and columns A and B Screws and washers from the carton. Place the scale platform and columns on a flat, level, and dry surface Remove the 4 column mounting screws and washers from the bottom of column B and set aside for assembly Column B...
  • Page 8 ASSEMBLY INSTRUCTIONS (CONTINUED) 3. Locate the load cell connector cable inside column A. While Column B Column A (top) holding column B, slide the (bottom) cable through the center opening at the top of column B. Push the cable all the way through Column B until the plug reaches the bottom of the column.
  • Page 9 ASSEMBLY INSTRUCTIONS (CONTINUED) 5. Place scale platform C on its Column B side and place Column B on its side, perpendicular to the platform. The front of the display head on Column A should be facing forward and above scale platform C. Column A 6.
  • Page 10 ASSEMBLY INSTRUCTIONS (CONTINUED) 7. Position and secure the connected columns A and B to scale platform C. Locate the 4 column mounting screws previously removed in step 2. Ensure that washers are located between the screw heads and the scale platform C screw holes.
  • Page 11: Powering The Scale

    POWERING THE SCALE Powering the Scale – Inserting Batteries 1. With a Phillips head screwdriver remove the screw from the battery compartment cover located on the back of the display. Remove the battery cover by pressing downward on the locking tab. Locking tab Display assembly Battery cover...
  • Page 12: Set Up Instructions

    SET UP INSTRUCTIONS To enter the Options mode, push and hold the ENTER button while pressing and releasing the button. Auto Off Function When powering the scale via the AC adapter, the display can be set to power off after a period of inactivity.
  • Page 13 SET UP INSTRUCTIONS (CONTINUED) Locking Unit of Measure (LB or KG) (499KL only, this feature is not available on 499KG) The weight measuring unit (pounds / LB or kilograms / KG) can be locked to display weight only in the selected unit of measure. The unit of measure lock default setting is “unlocked”. Follow this ®...
  • Page 14 SET UP INSTRUCTIONS (CONTINUED) pressing ZERO / TARE. Users will see dashes on the screen as the scale is finding zero. PU2 = E Enabling this feature means the scale will zero itself when it is powered on. After each use, the scale will clear the weight of the patient and set itself back to the ®...
  • Page 15: Operating Instructions

    OPERATING INSTRUCTIONS Keypad Function Description Turns scale on and off. Also used when entering Calibration or Options mode. Holds the displayed weight until the button is pressed again to HOLD/RLS/ release the value. Will recall the previously locked weight if pressed RECALL when nothing is on the scale.
  • Page 16 OPERATING INSTRUCTIONS (CONTINUED) Screen Icons Icon Description Indicates scale is being powered by batteries. Also indicates level of battery power remaining. Indicates scale is being powered by AC adapter. ZERO Displays along with dashes when scale is zeroing. Flashes during weighing process. Stays displayed while weight LOCK measurement is locked.
  • Page 17 OPERATING INSTRUCTIONS (CONTINUED) Weighing a Patient Note: When this scale is turned on, it will always default to the settings and units last used (LB or KG). 1. With all weight off the scale, press the button to turn on the scale. 2.
  • Page 18 OPERATING INSTRUCTIONS (CONTINUED) Weighing a Patient and Taring a Known Weight (Pre-Tare): Pre-Tare can be performed for items between 0.2 lb and 200 lb (0.1 kg and 91 kg). 1. With all weight off the scale, press the button to turn on the scale. 2.
  • Page 19: Assembly Instructions For Optional Height Rod (499Klrod)

    ASSEMBLY INSTRUCTIONS FOR OPTIONAL HEIGHT ROD (499KLROD) Note: These instructions are for mounting the optional height rod (499KLROD) to the 499KL/499KG scale. A height rod is not included with the 499KL/499KG scale and can be purchased separately as item # 499KLROD. Mounting 499KLROD 1.
  • Page 20: Height Measurement Instructions

    HEIGHT MEASUREMENT INSTRUCTIONS Note: These instructions are for measuring height using the optional height rod. A height rod is not included with the scale and can be purchased separately as item # 499KLROD. WARNING: Use caution when sliding height rod up and down. The rod should move smoothly.
  • Page 21 HEIGHT MEASUREMENT INSTRUCTIONS Height Rod Below 46 3/8” Height Rod Extended Above 46 3/8” Indicator Inches read: 69” + Inches read: 41” + Cm read: 176 cm + Cm read: 105 cm + 3/8” = 69 3/8” 3/8” = 41 3/8” 0.2 cm = 176.2 cm 0.2 cm = 105.2 cm Headpiece...
  • Page 22: Maintenance

    MAINTENANCE Maintenance The following pages provide instructions for maintenance, cleaning, calibrating, and troubleshooting of the scale. Maintenance operations other than those described in this manual should be performed by qualified service personnel. Caution: Before first use, or after long periods of non-use, check the scale for proper operation and function.
  • Page 23: Calibration

    CALIBRATION This scale has been factory calibrated, and does not require calibration prior to use. There are no requirements for calibration in the field; users should adhere to their institution’s calibration policies. If required, the scale can be calibrated. Calibration of your scale is performed using 80 kilograms (KG) or 200 pounds (LB), according to the units of measure used upon entering into the calibration path.
  • Page 24: Troubleshooting

    TROUBLESHOOTING Troubleshooting Before contacting service personnel, refer to the following instructions to check and to correct any failures. If display shows other error messages than displayed below, contact ® Health o meter Professional Scales Technical Support at 1-800-638-3722. Symptom Possible Cause Corrective Action Scale does not turn on 1.
  • Page 25: Warranty

    The original purchaser of the product must have proof of purchase to receive warranty service. Please save your invoice or receipt. Pelstar dealers or retail stores selling Pelstar products do not have the right to alter, or modify or any way change the terms and conditions of this warranty.
  • Page 26 Altura de la Cintura E112697 UL 60601-1 CSA C22.2 No. 601.1-M90 Instrucciones de Uso PELSTAR, LLC 9500 West 55th St. McCook, IL 60525-7110 USA www.homscales.com ® Health o meter Professional 499KL © Pelstar, LLC 2020 P/N UM499KL Rev 20201201...
  • Page 27 ® Health o meter Professional 499KL ® Gracias por comprar este producto Health o meter Professional. Lea con detenimiento este manual, y guárdelo para consulta rápida o capacitación. ÍNDICE Precauciones ......................3 Especificaciones ....................... 4 Certificaciones / Eliminación ................5 Instrucciones de Ensamblaje ..................
  • Page 28: Precauciones

    Contacte al Servicio al Cliente o a un servicio técnico especializado para que lo asista. En ningún caso, Pelstar, LLC será responsable de los daños o lesiones que surjan o estén relacionados con el ensamblaje, el uso o el uso indebido de sus productos.
  • Page 29: Especificaciones

    ESPECIFICACIONES General La balanza de columna digital a nivel de vista modelo 499KL/499KG utiliza tecnología computarizada muy sofisticada. Cada balanza de precisión está diseñada para mostrar medidas de peso exactas, confiables y repetibles. Además, cada balanza está diseñada para brindar al usuario características que hacen que el procedimiento de pesaje sea sencillo, rápido y práctico.
  • Page 30: Certificaciones / Eliminación

    (IMC), esta báscula está diseñada para conectarse a una computadora, monitor u otro ® dispositivo de datos electrónicos por medio de USB o la tecnología inalámbrica Pelstar . Esta conectividad ayuda a minimizar la Fuente de potenciales errores médicos causados al copiar y registrar los datos del paciente de manera incorrecta.
  • Page 31: Instrucciones De Ensamblaje

    INSTRUCCIONES DE ENSAMBLAJE Antes de ensamblar Cada balanza de columna digital a nivel de vista 499KL / 499KG se envía desarmada dentro de una caja. Revise la caja con detenimiento para localizar daños causados durante el envío antes de desempacar la balanza. Si encuentra algún daño, contacte de inmediato al transportista o a ®...
  • Page 32 INSTRUCCIONES DE ENSAMBLAJE (CONT.) 1. Retire tornillos arandelas de ensamblado de Tornillos y arandelas la base de la columna B y retire para su ensamblado en el paso 8. Columna B 2. Retire el soporte negro de la plataforma C. Lea la etiqueta de “Atención”...
  • Page 33 INSTRUCCIONES DE ENSAMBLAJE (CONT.) 3. Localice el cable conector de Columna B Columna A la celda de carga dentro de la (parte (parte columna A. Mientras sostiene superior) inferior) la columna B, deslice el cable a través de la apertura central en la parte superior de la columna Introduzca...
  • Page 34 INSTRUCCIONES DE ENSAMBLAJE (CONT.) 5. Coloque la plataforma C de la báscula acostada y también la Columna B, perpendicular a la plataforma. La parte frontal del cabezal de la pantalla en la Columna debe estar mirando hacia el frente y por encima de la plataforma C.
  • Page 35 INSTRUCCIONES DE ENSAMBLAJE (CONT.) 7. Posicione asegure columnas conectadas a la plataforma de la báscula C. Obtenga los 4 tornillos de montaje de la columna previamente retirados paso Asegúrese arandelas estén colocadas entre las cabezas de los tornillos y los orificios para tornillo de la plataforma C.
  • Page 36: Alimentando La Balanza

    ALIMENTANDO LA BALANZA Encendido de la Báscula – Insertar las Baterías 1. Con un destornillador de cruz, remueva el tornillo del compartimiento de las baterías ubicado en la parte posterior de la pantalla. Retire la cubierta de las baterías presionando hacia abajo la pestaña de sujeción.
  • Page 37: Instrucciones De Configuracion

    INSTRUCCIONES DE INSTALACIÓN Función Auto-Apagado Nota: El botón de Unidad está deshabilitado en modelos KG o KL activada o con la configuración “Unit View” deshabilitada. ® con la función Everlock Al hacer funcionar la báscula con el adaptador AC, la pantalla puede ser programada para apagarse luego de un periodo de inactividad.
  • Page 38 INSTRUCCIONES DE INSTALACIÓN (CONT) Bloqueo de la Unidad de Medición (LB o KG) (499KL solamente, esta característica no está disponible en el modelo 499KG) La unidad de medición de peso (libras / LB o kilogramos / KG) puede bloquearse para mostrar el peso en la unidad de medición seleccionada únicamente.
  • Page 39 INSTRUCCIONES DE INSTALACIÓN (CONT) Configuración de Puesta a Cero Automática Esta opción permite que el usuario seleccione si la báscula se va a ajustar a CERO cada vez que se encienda o si va a usar la lectura de cero de la última vez que fue puesta a cero. El ajuste de cero automático está...
  • Page 40: Instrucciones De Operación

    INSTRUCCIONES DE FUNCIONAMIENTO Teclado Función Descripción Enciende o apaga la báscula. También se usa cuando se ingresa a los modos de Calibración u Opciones. Retiene el peso mostrado hasta que el botón sea presionado otra HOLD/RLS/ vez para liberar el valor. Recordará el peso previamente retenido RECALL si está...
  • Page 41 INSTRUCCIONES DE FUNCIONAMIENTO (CONT.) Íconos en la Pantalla Icon Description Indica que la báscula está funcionando con baterías. También indica el nivel de batería restante. Indica que la báscula está funcionando con el adaptador AC. Exhibe junto con guiones cuando la báscula está poniéndose a ZERO cero.
  • Page 42 INSTRUCCIONES DE FUNCIONAMIENTO (CONT.) Nota: Cuando se enciende, la báscula siempre usará por defecto todos los ajustes y las unidades usadas por última vez (LB o KG). 1. Con la báscula libre de peso, presione el botón para encender la báscula. 2.
  • Page 43 INSTRUCCIONES DE FUNCIONAMIENTO (CONT.) Pesaje de un Paciente y Tara de un Peso Conocido (Pre-Tara) La Pre-Tara puede realizarse para objetos entre 0.2 lb y 200 lb (0.1 kg y 91 kg). 1. Con la báscula libre de peso, presione el botón para encender la báscula.
  • Page 44: Instrucciones De Ensamblaje Para 499Klrod Opcional

    INSTRUCCIONES DE ENSAMBLAJE PARA 499KLROD OPCIONAL (499KLROD) Nota: Estas instrucciones son para instalar el tallimetro opcional (499KLROD) a la balanza 499KL / 499KG. El tallimetro (499KLROD) no está incluido en la balanza y debe adquirirse por separado. Montaje del 499KLROD 1.
  • Page 45: Instrucciones Para La Medición De La Altura

    INSTRUCCIONES PARA LA MEDICIÓN DE LA ALTURA Nota: Estas instrucciones son para medir la altura usando la barra de altura opcional. Una varilla de altura no está incluida en la escala 499KL/499KG y se puede comprar por separado como artículo # 499KLROD. ADVERTENCIAS: Tenga cuidado al deslizar la barra de altura hacia arriba y hacia abajo.
  • Page 46 INSTRUCCIONES PARA LA MEDICIÓN DE LA ALTURA Estadímetro menor a 46 3/8" Estadímetro extendido mayor a 46 3/8" Indicador Lectura en pulgadas: Lectura en centímetros: 105 Lectura en centímetros: 176 Lectura en pulgadas: 69” + 3/8” = 69 3/8” cm + 0.2 cm = 176.2 cm 41”...
  • Page 47: Mantenimiento

    MANTENIMIENTO Mantenimiento Las páginas siguientes proporcionan las instrucciones para el mantenimiento, la limpieza, la calibración, y la resolución de problemas de la báscula. Las operaciones de mantenimiento fuera de las descritas en este manual se deben realizar por personal de servicio calificado. Precaución: Antes del primer uso, o después de períodos largos de falta de uso, revise la operación y funciones apropiadas de la báscula.
  • Page 48: Calibración

    CALIBRACIÓN Esta báscula ha sido calibrada de fábrica y no requiere calibración previo a su uso. No hay requerimientos para la calibración en campo; los usuarios deben ajustarse a las políticas de calibración de su organización. Si es necesario, la báscula puede ser calibrada.
  • Page 49 RESOLUCIÓN DE PROBLEMAS Resolución de Problemas Antes de entrar en contacto con el personal de servicio, consulte las instrucciones siguientes para comprobar y corregir cualquier falla. Si la pantalla muestra mensajes de error distintos a ® los mostrados abajo, contacte a: Soporte Técnico de Health o meter Professional Scales al 1- 800-638-3722.
  • Page 50: Garantía

    El comprador original del producto debe tener un comprobante de compra para recibir el servicio de garantía. Por favor guarde su factura o recibo. Los distribuidores de Pelstar o las tiendas al menudeo que venden los productos de Pelstar no tienen el derecho de alterar o modificar o en forma alguna cambiar los términos y condiciones de esta garantía.

This manual is also suitable for: