Mounting On Pipe Socket - Xylem mjk pHix Compact Manual

Hide thumbs Also See for mjk pHix Compact:
Table of Contents

Advertisement

Because of the liquid inside the electrode, pHix® Compact should always be
mounted in angles less than 75° from vertical.
pHix® Compact can be screwed into a pipe stub with 2" outer thread. This can
happen by means of an union (item no. 521409) placed over the housing
flange after removal of the switch ring.
Pull off the switch ring.
The switch ring is only used for buffer adjustment (calibration), but it is also
possible to perform a
ring.

Mounting on pipe socket

In order to ensure a tight seal and correct function of pHix® Compact, the
following should be complied with:
1.
The pipe socket must be flushed and free of burrs, etc.
2.
The pipe socket must have a length so that the tip of the electrode is
immersed min. 10 mm into the liquid.
Mounting / Mechanical mounting
Put on the union.
buffer adjustment
Put on the switch ring if
without using the switch
 (see page 24)
pHix® Compact
EN 4.10  Version 2007
needed.
17

Advertisement

Table of Contents
loading

Table of Contents