pH and temperature transmitter, item no. 203113
Measuring range: 0-14 pH = 4-20 mA, 0-point = 4,6 pH
Temperature: 32-122° F = 4-20 mA
Redox transmitter, item no. 203121
Measuring range: -1000 to +1000 mV = 4-20 mA
ALL OTHERS:
pH and temperature transmitter, item no. 203110:
Measuring range: 0-14 pH = 4-20 mA, 0-point = 4,6 pH
pH transmitter, item no. 203112:
Measuring range: 0-14 pH = 4-20 mA, 0-point = 4,6 pH
Temperature: 0-50° C = 4-20 mA
Redox transmitter, item no. 203120:
Measuring range: -1000 to +1000 mV = 4-20 mA
HART® commands
All commands and settings can be read and set by means of HART®
commands. Ask for separate documentation regarding HART® commands.
Appendencies / HART® commands
pHix® Compact