pHix® Compact
EN 4.10  Version 2007

Repair

Repair must only made by MJK or by a service representative approved by
MJK.

Explosion hazardous areas

Note! pHix® Compact must not be installed in explosion hazardous areas!

Construction

pHix® Compact is a loop powered 4-20 mA 2-wire transmitter with power
supply and measuring signal transmitted over the same two wires.
pHix® Compact also has HART® communication capabilities via the mA wires
where both the primary measuring signal (pH or redox) and the secondary
measurement signal can be read. HART® communication gives the possibility
to configure nearly all parameters like measuring ranges, alarm limits, electrode
parameters and also commands for buffer adjustment.
HART® communication requires that pHix® Compact is connected to a HART®
communicator or a PLC with HART® interface.
pHix® Compact is delivered as standard for pH or redox measurement, but can
also be delivered for other types of measurement and non-standard settings.

Product identification

Check that the item(s) delivered corresponds to the ordered item(s). The item
number is printed on a label that is sticked onto the packing. Shown in the
following, is the label for a delivery including a pHix® Compact transmitter with
pH electrode with zero point at pH 4,6: