Xl-80Q Operation With Rcu10 - Renishaw XL-80 Manual

Laser system
Hide thumbs Also See for XL-80:
Table of Contents

Advertisement

XL-80 Hardware
XL-80 Hardware
Linear
XL-80 Applications
Angular

XL-80Q operation with RCU10

XL-80Q set-up:
XL-80Q DIP switch 2 must be set for appropriate quadrature resolution
"ON" = ≈10 nm resolution (λ/64)
"OFF" = ≈80 nm resolution (λ/8)
XL-80Q to RCU10 cable
Below is a summary of the wiring required between an XL-80Q (Auxiliary I/O
connector) and an RCU10 (15-pin D-type male connector)�
NOTE: A cable with twisted pairs and overall shield is recommended for the
digital quadrature interface between the RCU10 and the XL-80Q�
1st pair
2nd pair
3rd pair
59
XL80 laser system
Straightness
A Quad and /A Quad
B Quad and /B Quad
Error and /Error
XL-80Q
Pin number
Signal
2
0V
22
/Alarm
23
Alarm
7
/B Quad
20
/A Quad
8
B Quad
21
A Quad
4 and 16
External reset switch
RCU10
Signal
Pin Number
0V
2
/Error
3
Error
11
/B Quad
5
/A Quad
6
B Quad
13
A Quad
14
www�renishaw�com/xl80