Situation
Generator is on and the
accessory is activated, but
the generator does not deliver
output.
7-8
Possible Cause
Malfunctioning footswitch or
handswitching instrument
Power set too low
Display board malfunction
Main board malfunction
Relay board malfunction
NEM alarm
Recommended Action
1. Turn off the generator. Check and correct all
accessory connections.
2. Turn on the generator
3. Replace the accessory if it continues to
malfunction.
Increase the power setting.
Replace the display board.
Replace main board.
Replace main board.
1. Check/connect the Return Electrode connection
to the patient and connection to the generator.
2. Replace Return Electrode