Voltage and current are sensed during power output.
CONTROLS AND INDICATORS
The Aaron 2250 controls and indicators are listed below:
• MEMBRANE SWITCHES Toggle between modes.
• DISPLAYS Three separate displays (Cut, Coag, & Bipolar) indicate the output power in watts.
• MODE INDICATORS Green LEDs indicate the present mode of the unit.
• POWER CONTROL SWITCHES These membrane switches adjust the output power for each mode.
• POWER SWITCH A double pole single throw switch that snaps into the front bezel. This switch supplies the AC
mains current to the generator.
AARON 2250 CONTROL SIGNAL INPUTS AND OUTPUTS
CUT/COAG
RF_DRV1
RF_DRV2
DAMP
The following table lists the important input and output signals. From a troubleshooting standpoint, the absence
(and presence) of these signals will assist in isolating problems.
Ser vice Guide • Aaron
This is an output signal that defines the modulation signal to the output stage.
This is an output signal from the system logic that generates the driving signal for the
output stage.
This is an output signal from the system logic that generates the driving signal for the
output stage.
This is an output signal from the system logic to control the damping of the output.
4-5