Bovie AARON 2250 Service Manual page 48

Table of Contents

Advertisement

A/D1_REF_V_DAC_DATA
A/D1_REF_V_DAC_SCLK
A/D1_REF_V_DAC_SYNC
A/D2_REF_V_DAC_DATA
A/D2_REF_V_DAC_SCLK
A/D2_REF_V_DAC_SYNC
BURN_TST
AUX1_JMP
TEST_JMP
I_TRANS_1
I_TRANS_2
I_TRANS_3
V_TRANS_1
V_TRANS_2
V_TRANS_3
L_TRANS_1
L_TRANS_2
L_TRANS_3
TEMP_1
TEMP_2
TEMP_3
VB_TRANS_1
VB_TRANS_2
VB_TRANS_3
IB_TRANS_1
IB_TRANS_2
IB_TRANS_3
4-10
This is a communication line used by the system logic to control the sensing A/D's.
This is a communication line used by the system logic to control the sensing A/D's.
This is a communication line used by the system logic to control the sensing A/D's.
This is a communication line used by the system logic to control the sensing A/D's.
This is a communication line used by the system logic to control the sensing A/D's.
This is a communication line used by the system logic to control the sensing A/D's.
This is an input signal to the control logic for toggling the unit into automatic burn-in routine.
Reserved
Reserved
This is an input signal from the current sensor to the control logic. Used to monitor the
output current of the unit.
This is an input signal from the current sensor to the control logic. Used to monitor the
output current of the unit.
This is an input signal from the current sensor to the control logic. Used to monitor the
output current of the unit.
This is an input signal from the voltage sensor to the control logic. Used to monitor the
output voltage of the unit.
This is an input signal from the voltage sensor to the control logic. Used to monitor the
output voltage of the unit.
This is an input signal from the voltage sensor to the control logic. Used to monitor the
output voltage of the unit.
This is an input signal from the impedance sensor to the control logic. Used to monitor the
output voltage of the unit.
This is an input signal from the impedance sensor to the control logic. Used to monitor the
output voltage of the unit.
This is an input signal from the impedance sensor to the control logic. Used to monitor the
output voltage of the unit.
This is an input signal from the temperature circuit to the control logic. Used to monitor
internal temperature of the unit.
This is an input signal from the temperature circuit to the control logic. Used to monitor
internal temperature of the unit.
This is an input signal from the temperature circuit to the control logic. Used to monitor
internal temperature of the unit.
This is an input signal from the backup voltage sensor to the control logic. Used to monitor
the output voltage of the unit.
This is an input signal from the backup voltage sensor to the control logic. Used to monitor
the output voltage of the unit.
This is an input signal from the backup voltage sensor to the control logic. Used to monitor
the output voltage of the unit.
This is an input signal from the backup current sensor to the control logic. Used to monitor
the output current of the unit.
This is an input signal from the backup current sensor to the control logic Used to monitor
the output current of the unit.
This is an input signal from the backup current sensor to the control logic. Used to monitor
the output current of the unit.