4. Check for a secure fit by inserting the footswitch connectors into footswitch receptacles.
If the footswitch receptacles are damaged, replace the footswitch connector assembly.
5. Check the bipolar receptacle on the front of the unit for obstruction or damage.
6. Insert a bipolar cable into the bipolar receptacle on the front of the unit. Verify a secure fit.
If the bipolar receptacle is damaged, replace the bipolar connector assembly.
7. Check the monopolar handpiece receptacle on the front of the unit for obstruction or damage.
8. Insert a handswitching pencil into the monopolar handpiece receptacle on the front of the unit.
Verify a secure fit.
If the monopolar handpiece receptacle is damaged, replace the monopolar handpiece assembly.
9. Check the monopolar foot controlled receptacle on the front of the unit for obstruction or damage.
10. Insert a monopolar foot controlled handpiece into the monopolar foot control receptacle on the front of the unit.
Verify a secure fit.
If the monopolar foot controlled receptacle is damaged, replace the connector assembly.
11. Check the Return Electrode receptacle on the front of the unit for a broken pin or an obstruction.
12. Insert a return electrode cable into the return electrode receptacle. Verify a secure fit.
If the return electrode receptacle on the front of the unit is damaged, replace the return electrode cable assembly.

Inspecting Internal Components

CAUTIONS:
The generator contains electrostatic-sensitive (ESS) components. When repairing the generator,
work at a static-control workstation.
Wear a grounding strap when handling electrostatic-sensitive components.
Handle circuit boards by their nonconductive edges.
Use an anti-static container for transport of electrostatic-sensitive components and circuit boards.
To inspect the internal components, follow this procedure:
1. Remove the six screws that secure the cover to the chassis.
2. Lift the cover off the chassis. Save the cover and screws for later reinstallation.
3. Visually inspect and verify that all connectors are firmly seated.
4. Inspect each board for damaged components, wires, cracks and corrosion.
5. Reinstall the cover by positioning the cover over the chassis, and securing the four 6 - 32 x
tom sides of the chassis and the two 6 - 32 x
Ser vice Guide • Aaron 2250
1
" screws to the lower front chassis.