PERFORMANCE CHARACTERISTICS
Input Power
Input Voltage
Mains line frequency range (nominal):
Power consumption:
Fuses (two):
Duty Cycle
Under maximum power settings and rated load conditions (Cut, 200 watt @ 300 ohm load), the generator is suitable
for activation times of 10 seconds ON followed by 30 seconds OFF for one hour.
The internal temperature of the unit is continuously monitored. If the temperature rises above 85
sound and output power will be deactivated.
Dimensions and Weight
Width
31.1 cm (12.25 in.)
Height
15.3 cm (6.0 in.)
Operating Parameters
Ambient temperature range
Relative humidity
Atmospheric pressure
Warm-up time
Transport and Storage
Ambient temperature range
Relative humidity
Atmospheric pressure
Generator should fit on all standard Carts for monopolar generators. The device should be stored and used in a
room temperature of approximately 77
3-2
100-240 VAC
50 / 60 Hz
500 VA
6.3 A (slow blow)
Depth
Weight
10° to 40° C (50° to 104° F)
30% to 75%, non-condensing
700 to 1060 millibars
If transported or stored at temperatures outside the operating temperature range,
allow one hour for the generator to reach room temperature before use.
-34˚ to 65˚ C (-29˚ to 149˚ F)
0% to 75%, non-condensing
500 hPa to 1060 hPa
0
F/25
0
C.
0
C, the alarm will
41.3 cm (16.25 in.)
< 8.75 kg (< 19 lbs)
Aaron Medical