PERFORMANCE CHARACTERISTICS

Input Power

Input Voltage
Mains line frequency range (nominal):
Power consumption:
Fuses (two):

Duty Cycle

Under maximum power settings and rated load conditions (Cut, 200 watt @ 300 ohm load), the generator is suitable
for activation times of 10 seconds ON followed by 30 seconds OFF for one hour.
The internal temperature of the unit is continuously monitored. If the temperature rises above 85
sound and output power will be deactivated.

Dimensions and Weight

Width
31.1 cm (12.25 in.)
Height
15.3 cm (6.0 in.)

Operating Parameters

Ambient temperature range
Relative humidity
Atmospheric pressure
Warm-up time

Transport and Storage

Ambient temperature range
Relative humidity
Atmospheric pressure
Generator should fit on all standard Carts for monopolar generators. The device should be stored and used in a
room temperature of approximately 77
3-2
100-240 VAC
50 / 60 Hz
500 VA
6.3 A (slow blow)
Depth
Weight
10° to 40° C (50° to 104° F)
30% to 75%, non-condensing
700 to 1060 millibars
If transported or stored at temperatures outside the operating temperature range,
allow one hour for the generator to reach room temperature before use.
-34˚ to 65˚ C (-29˚ to 149˚ F)
0% to 75%, non-condensing
500 hPa to 1060 hPa
0
F/25
0
C.
0
C, the alarm will
41.3 cm (16.25 in.)
< 8.75 kg (< 19 lbs)