FUSE REPLACEMENT ON THE MAIN PCB

Additional fuses for the unit reside on the Main PCB. Fuse values are indicated in the table below.
To replace the fuses, follow this procedure:
1. Unplug the power cord from the wall outlet.
2. Remove the power cord from the Power Cable Receptacle on the rear panel.
3. Remove the six screws that secure the cover panel to the unit.
4. Lift off the top cover panel.
5. Remove the two fuses using a fuse pulling tool.
6. Replace the fuses with the same values as listed below.
7. Replace and secure the cover panel.
Figure 6 – 2
Fuse location

Main PCB Fuse Information

VAC
AMPS
Size
Type
Ser vice Guide • Aaron 2250
FS1
250 VAC
6.3 A
5 x 20 mm
Slow Blow
FS2
250 VAC
0.5 A
5 x 20 mm
Slow Blow
6-3