PERFORMANCE CHARACTERISTICS
Input Power
100 – 240 VAC
Mains line frequency range (nominal): 50 – 60 Hz
Power consumption: 270 VA
Fuses (two): 3.15A (Slow Blow)
Duty Cycle
Under maximum power settings and rated load conditions (Cut, 120 watt @ 500 ohm load), the generator is suitable
for activation times of 10 seconds on, 30 seconds off for one hour.
The internal temperature of the unit is continuously monitored. If the temperature rises above 850 C, the alarm will
sound and output power will be deactivated.
Dimensions and Weight
Width
Height
Operating Parameters
Ambient temperature range
Relative humidity
Atmospheric pressure
Warm-up time
Transport
Ambient temperature range
Relative humidity
Atmospheric pressure
Storage
Ambient temperature range
Relative humidity
Atmospheric pressure
Service Guide • Specialist | PRO
26 cm (10.25 in.)
15.2 cm (6 in.)
10° to 40° C (50° to 104° F)
30% to 75%, non-condensing
70kPa to 106kPa
If transported or stored at temperatures outside the operating temperature range,
allow one hour for the generator to reach room temperature before use.
-40° to +70° C
10% to 100%, including condensation
50kPa to 106kPa
10° to 30° C (68° to 86° F)
10% to 75%, non-condensing
50kPa to 106kPa
Depth
30.5 cm (12 in.)
Weight
< 4 kg (< 9 lbs)
27