MAINTENANCE OF THE PRODUCT

1. Cleaning
■ Angle Handpiece
① Rotate the head in a tap water about 10 seconds with maximum speed to remove the remains
(blood, physiologic saline solution, etc.) and rinse.
※ Do not put tap water in the motor.
② Disconnect the angle handpiece to the motor.
③ Remove contaminants on the angle handpiece using a nylon brush after soaking for 20 minutes
in Enzymatic Detergent.
④ Rinse the angle handpiece using the tap water for 3 minutes to remove Enzymatic Detergent.
⑤ Wash the angle handpiece for 10 minutes in the ultrasonic cleaner containing Enzymatic Detergent.
⑥ Rinse the angle handpiece for 3 minutes in the ultrasonic cleaner containing purified water repeatedly
subjected 3 times. (Change the purified water each time.)
⑦ Wipe using dust-absorbent after the product soaked in alcohol(70%) for 30 seconds.
■ Motor
Wipe the surface of motor with alcohol swab or paper towel after removing the dirt.
※ Pay attention to not enter the water into the motor.
- Proteolytic enzyme detergents are recommended as Enzymatic Detergent.
- It is recommended to use ethanol for alcohol (70%).
2. Sterilization
Sterilization Product List
· Angle handpiece
· Motor (Assembling with motor cap)
Sterilization Method
① Dry the applicable products completely after cleaning.
② Put the applicable products in a sterile bag and seal tightly.
※ Make sure to assemble the motor cap for autoclave on the motor.
※ Do not disassemble the motor and motor cord when sterilizing.
③ Put the sterile bag in the sterilizer and sterilization for 4 minutes with 132℃, and let it dry for 30 minutes.
Sterilization
Moist heat sterilization
- This product is designed to withstand 250 times of re-sterilization process.
- This product recommends the moist heat sterilization (autoclave), other sterile methods are
not guaranteed the effectiveness.
- Make sure to use the sterilization bag and seal tightly.
- Make sure to dry the applicable products before sterilization.
- Motor could be damaged in sterilization because of the moisture inflow or else, so should be careful to handle it.
- Sterilizing the angle handpiece with blood or other remains may cause the breakage of the angle handpiece.
Therefore, it is necessary to clean the angle handpiece before sterilization.
- Irrigation tube and Y-tube are disposable. Do not re-use.
- Keep dry a connector and socket of the control box.
- Before using, dry the product completely.
Temperature
132℃
Time
4 minutes
30 minutes
132
Dry
30