Insufficient disk
space.
Please connect the
camera.
Unsuccessful
connection to the
printer.
Please confirm
whetherPDFCreator
0.9.6.0 is installed.
WorkStation of
Colposcope stops
working.
The disk space is
insufficient.
The camera is not
switched on.
The USB-to-COM driver
has not been installed.
Select the incorrect
printer in the print setup.
The printer drive is not
installed correctly.
No connection to the
printer.
No PDFCreator 0.9.6.0 is
installed.
The software breaks
down.
operating system is used, open the X64
folder, right-click
installUSBCard.batand choose Run as
administrator.
C6A HD: Open the software installation
package and find hd, double-click and
run DRIVER.QP0204_YS.exe.
Release some disk space.
Connect equipment forming the system
as required and switch on the camera.
1. Connect the camera to the computer
and switch on the camera.
2. Open the software installation
package and find the
usbdriver\USBToCOMdirectory. If the
32-bit operating system is used, open
the X32 folder, right-click
installUSBtoCOM.bat and choose Run
as administrator. If the 64-bit
operating system is used, open the X64
folder, right-click
installUSBtoCOM.bat and choose Run
as administrator.
Open the print setup page and select the
correct printer.
Reinstall the printer drive.
Connect to the printer.
Install PDFCreator 0.9.6.0.
Restart the software.
- 97 -
Troubleshooting