To prevent injury to people and damage to property, INDICATIONS FOR USE: please heed relevant warnings and remarks. They are The ADU-17A is a mobile self-contained dental system marked as follows: that is used for endodontic and general dentistry applications.
SAFETY PRECAUTIONS: Your new Aseptico Express Air Portable Dental System is the finest portable electric dental system available. The ADU-17A Aseptico accepts no liability for direct or consequential injury or features a handpiece control, 3-way air/water syringe, HVE and damage resulting from improper use, arising in particular through...
SETTING UP THE UNIT: FIG. 1 - ADU-17A Express Air HANDPIECE 3-WAY AIR/WATER SYRINGE HIGH VOLUME VACUUM (HVE) CLIP-ON INSTRUMENT HOLDER LOW VOLUME VACUUM (SALIVA EJECTOR) CASE HANDLE WASTE TANK TIE- DOWN RESEVOIR CONNECTOR WATER SUPPLY BOTTLE WASTE TANK COOLANT ON/OFF...
Page 5
Fig. 2a Fig. 2b Fig. 3 Clip-On Holder Handle 120V Indicator Saliva Ejector 3-Way 240V Indicator Syringe Voltage Selector Module Handpiece Attach a sterile handpiece to the ISO Type B (4-Line) Fig. 2c Fig. 2d handpiece connector and place the handpiece in Holder (Fig.
TURN CONNECTOR 1/4 TURN CLOCKWISE Fig. 5 ON/OFF SWITCH: VACUUM LINE WITH QUICK Turns the ADU-17A system On or Off. DISCONNECT AIR FOOT CONTROL: The handpiece only operates when the foot pedal is depressed. When the foot pedal is not being used, the vacuum and syringe remain operational.
CAUTION: CAVITRON WATER OUTLET: Some new ADU-17A compressor motors might require Provides water via standard 1/4” quick disconnect to a 2-week “break-in” when first used. As a precaution against stalling the compressor or possibly blowing the independent Cavitron or scaler units requiring a water supply.
CAUTION: Use only NON-foaming cleansers in the vaccum lines. SYRINGE TIPS The ADU-17A unit features a three-way air/water syringe with quick-change autoclavable tips. To remove a tip, press on the locking collar surrounding the tip socket and pull...
TROUBLESHOOTING: EXTENSION CORD RECOMMENDA- TIONS: Unit will not start Use only a 3-wire extension cord that has a 3-blade grounding Check system’s power connection. plug. Connect extension cord plug to a matching 3-slot receptacle. Ensure that the extension cord is in good condition Turn On/Off switch ON for operation.
Note: For reliable grounding, connect toreceptacle marked “Hospital Grade”. IMPORTANT NOTE: When running the ADU-17A unit at 230VAC 50Hz, expect approximately 17% less vacuum and pressure volume due to the slower turning of the compressor. This device has been tested and found to comply with the emissions requirements CAUTION: of IEC 60601-1-2:2001-09.