USE
While the brush and the endoscope are submerged in a cleaning
solution, advance the leading tip of the DART
into the endoscope channel opening until the leading brush head
emerges out the other end of the channel. Use short strokes to avoid
kinking. Hold the leading tip of the brush and pull brush until trailing tail
of the brush clears the channel exit (see Figure 9 & 11). With the
brush still submerged in the detergent solution, scrub off all debris on
the bristles. Perform at least two passes through each channel or until
no more visible debris upon inspection of brush.
4
Figure 8 Direction of use
Figure 9 Cleaning suction and instrument channels
Figure 10 Cleaning instrument channel
®
Start with leading tip of
DART
2.0 cm
Hold leading tip and pull until
tail of brush exit channel
Hold leading tip and pull until
tail of brush exit channel
brush (see Figure 8)
®
brush
Shaft