External Trigger; Trigger Button - Protek B8000FD Series Operator's Manual

Dds function generator
Table of Contents

Advertisement

FREQ1000
VOLT100
DUTY50
OFFSET0
Fig 28
Each instruction must not exceed 15 characters.
Every instruction must end with ";" or with a return character
to indicate that the instruction set is sent out. The ASCII code of
a new line character is 10, but it is indicated by "\n" in C
language. If there is no ending character in the instruction set,
the instrument may "lock up" and be restarted.
There should be sufficient time between two instruction so as
to allow enough time for the instrument to finish the current
operation.

6 External Trigger

6.1 Trigger Button

When in the SWEEP, FSK, ASK and BURST modes it may be
at times desirable to execute one cycle at a time. The B8000FD
series of DDS generators have single cycle and external trigger
capability.
Pressing "Trig" key once, the instrument enters into
triggered mode. The LCD will display "trig" on the bottom right
Sets the frequency to 1000Hz
The range is from 0.001 to15,000,000
Sets the amplitude to 100mV, the
range is 1mV to 10,000mV
Sets the pulse ratio to 50% duty
ratio. The duty ratio range is 20% to
80%
Set the DC of set offset to 0%. The
DC offset, range is -100% to +100%
24