MAIN SCREEN

5.3
The main screen displays:
Name of the analysis program (Silica Analyzer)
Login **** window (two levels of password); pressing on **** the user can input the 4
number password
Buttons that allow access the sub-menus (Run, Display, Program, Service and ?)
Value (ppm or ppb) of the last analysis
Alarms / relays
Active step of the analysis in progress
Reagent(s) consumption (%)
STAND-BY The analyzer is waiting for a user's command (for example when the analyzer
is switched on after a shutdown), after single cycle analysis, or when the
operator has forced a stop to stand-by at the end of the current online / single
cycle by pressing the button previously activated a second time.
SINGLE
The analyzer is performing a single analysis cycle. As soon as the cycle is
CYCLE
completed the analyzer will go into the stand-by condition.
ON-LINE
The analyzer is performing a continuous cyclic analysis based on the steps
set in the PROGRAM menu (Analysis Cycle) and started when the START
ON LINE button has been pressed. As soon as the analyzer completes an
analysis cycle, it will restart with a new analysis. This condition is identified as
the online condition.
MANUAL
The analyzer has been forced to stop. A message in the main menu shows
the "manual stop" and the fault alarm is initialized.
STOP
RUN
This menu allows the user to put the analyzer into online measurement mode
or run a single analysis cycle on stream 1 or stream 2. The user can also
stop the analyzer with the emergency stop button.
DISPLAY
This menu displays all analysis data, the active step of the current analysis
cycle with the time counter, the sensor's signal trend graph, and allows
access to the manual step functions.
PROGRAM First level service password required. This menu allows access to the
calibration menu, enabling and disabling the loss of sample alarm, enabling
the 4-20 mA output, and settings of the analyzer.
SERVICE
Second level admin password required. This menu allows access to the
analysis parameter, time, blank value, and configuration of the 4-20 mA
output, led light source, and reference value.
DATA LOG This menu shows the data log.
Waltron User Manual 101-016-B.3
3041 Silica Analyzer

Table 5.3: Analyzer operating statuses.

ANALYZER OPERATING STATUS

Table 5.4: Main menu button descriptions.

MAIN MENU BUTTONS
User Interface
32