Overview; Specifications; Performance; Sample Operating Conditions - Waltron 3041 User Manual

Silica analyzer
Table of Contents

Advertisement

Overview

1 OVERVIEW

SPECIFICATIONS

1.1

PERFORMANCE

1.1.1
Sensor Classification:
Application:
Power Supply:
Humidity:
Ambient Temperature Range:
Range:
Accuracy:
Unit Dimensions:
Positioning and Installation
Details:
Response Time (approx.):
Alarms:
Reproducibility:
Degree of Protection:
Required Maintenance:
Reagent Consumption
(approx.):
Materials in Contact with
Sample:
CE Certification (on request):
Sample Conditioning
Requirements:
Weight:

SAMPLE OPERATING CONDITIONS

1.1.2
Temperature Range:
Inlet Sample:
Outlet Sample:
Turbidity:
Color:
PH:
9
Colorimetric with glass flowcell and LED source
Demineralized, boiler, potable, surface and waste
water
110-220Vac, 50-60 Hz 80 VA
Up to 90% not condensable
10-45º C analyzer (50-113º F)
0 – 500ppb, 0 – 5ppm, 0 – 150ppm*, 0 – 300ppm*
*By Auto Dilution
+/– 1 ppb or +/– 2% of reading, whichever is greater
Height=23.6"(60 cm), Width=15"(38 cm),
Depth=8.3"(21 cm)
Wall mounting or with bench support, panel mount
optional
12min, batch process, programmable cycle frequency
2 or 4 configurable alarm relays
+/– 1 ppb or +/– 2% of reading, whichever is greater
IP55 (NEMA 4): Wet section, IP65 (NEMA 4x):
Electronics
Monthly replenish reagents, Quarterly replace tubing
3 reagents, 2.5L each per month (optional 7.5L for 3
months)
Glass, Silicone, Plexiglass, Stainless Steel AISI 316
Meets low voltage and low electromagnetic
compatibility directives
Filtering between 10 and 60 micron, depending on the
matrix, needed only to avoid clogging
17 kg
5-55º C (41-131º F)
100–500 ml/min, atmospheric, flexible tubing I.D. ¼" (6 mm)
Atmospheric, waste tubing O.D. 3/8" (9 mm)
Not applicable, sample blank correction
Not applicable, sample blank correction
3-12
Waltron User Manual 101-016-B.3
3041 Silica Analyzer