9 TROUBLESHOOTING

For troubleshooting tips and solutions, refer to the HELP Menu. The HELP Menu can be found
through the user interface of the analyzer and in Appendix B of this manual.
Alarm
Message
Reagent
<10%
Zero Failure
Loss of
Sample 1
Loss of
Sample 2
Emergency
Stop
Cal Error
Result Alarm
Waltron User Manual 101-016-B.3
3041 Silica Analyzer

Table 9.1: Alarm descriptions.

Alarm Description
One or more reagent
bottles is low
The Reference Failed
The sample to stream 1
is lost
The sample to stream 2
is lost
Someone felt it was
necessary to press the
Emergency Stop button
The Calibration Factor is
out of Range.
Measurement is greater
than the % on the
Settings screen
Action Required
Check the level of the reagent bottles and
refill as necessary
Check the water level in the sample cell
Check the flow of Sample #1
Check the flow of Sample #2
Dual stream analyzers only
Determine why Emergency Stop was
pressed, correct problem and then press the
RESET button on the RUN window.
1. Check the level of the reagent bottles,
2. Make sure reagent pumps are primed
3. Check reagent pump dosage
4. Check Cal Sample level
5. Verify analyzer is draining properly
The sample is out of range. Corrective
action required.
Troubleshooting
50