9065 Luminescent Dissolved Oxygen Analyzer Calibration Kit
Quick Reference Manual
104-001-A.7 Released 06/14/2019

Advertisement

loading

Summary of Contents for Waltron 9065

  • Page 1 9065 Luminescent Dissolved Oxygen Analyzer Calibration Kit Quick Reference Manual 104-001-A.7 Released 06/14/2019...
  • Page 2: Warranty Agreement

    WALTRON CUSTOMER COMMITMENT This quick reference manual is a technical guide to aid the customer in the calibration of the Waltron 9065 Dissolved Oxygen Analyzer using the 9065 calibration kit. Waltron provides continuous product improvement and reserves the right to make any modifications to the information contained herein without notice.
  • Page 3: Introduction And Setup

    Make sure you have all items A calibration beaker is provided necessary to complete calibration. with each 9065 analyzer order. Inventory your calibration kit Contact Waltron for additional (Calibration Kit p/n K6000-101). beakers p/n K5000-039.
  • Page 4 Step Description Key Points Disconnect the plumbing from the flow cell. Remove the flow cell. DO NOT wipe dry the black Allow the sensor to air dry and to sensor spot. Touching the allow the sensor and the sensor may damage the calibration kit to equalize to the delicate coating.
  • Page 5 Step Description Key Points CONNECT THE NITROGEN CALIBRATION GAS Attach M5 barb fitting (provided) to calibration beaker inlet. Ensure the provided O-Ring is The inlet is the side that is closer installed on the fitting to to the flange than the other provide a seal against leaks fitting.
  • Page 6 Step Description Key Points Remove the protective cap from the provided nitrogen tank. Install the gas regulator on nitrogen gas tank. Install one end of the provided tubing on to gas regulator. 104-001-A.7 Released 06/14/2019...
  • Page 7 Step Description Key Points Install the other end of the provided tubing on to the barb fitting on the calibration beaker. The provided fixed flow Open the regulator all the way to regulator will supply the allow flow of Nitrogen gas. correct flow rate.
  • Page 8 If either PHASE or AMPLITUDE do Contact Waltron for a not fall within these ranges then replacement sensor spot p/n the sensor spot needs K5000-010. Refer to 9065 User replacement. manual for installation. 104-001-A.7 Released 06/14/2019...
  • Page 9 Step Description Key Points Using the Arrow buttons select O2 AUTO. Press ENTER button. In the next screen select O2 LOW. Press ENTER button. Select 000000 ppb. Press ENTER button. Calibration will begin. Wait approximately 2 minutes. Do not disturb the analyzer or the calibration equipment.
  • Page 10 If calibration is not successful, analyzer will show an alert for make sure the sensor spot is successful calilbration. completely dry and try again. Contact Waltron technical Press any button to clear this support if you are still alert. unsuccessful.
  • Page 11 Step Description Key Points After calibration is successfully stored, cycle the power to reboot the analyzer by disconnecting and then reconnecting the power cord (connector at the bottom of the transmitter which is furthest to the left). Make sure the valve on the gas regulator is closed. Remove the gas regulator from the gas cylinder.