RUN MENU

5.5

Figure 5.3: User interface after pressing the RUN Menu.

START ON-LINE:
This button puts the analyzer into continuous online operation. The analyzer will perform a
continuous cyclic analysis based on the steps set in the PROGRAM menu (Analysis Cycle).
This condition is identified as online and ONLINE is displayed on the button. Pressing the
button again will cause the analyzer to finish the cycle in progress and put the analyzer in
stand-by mode.
CH 1:
This command will perform a single analysis cycle on stream 1. At the end of the cycle, the
analyzer will go into stand-by mode and wait for the user's next command.
CH 2 (available on dual stream units only):
This command will perform a single analysis cycle on stream 2. At the end of the cycle, the
analyzer will go into stand-by mode and wait for the user's next command.
EXTRA:
This command will perform a cycle based on the programming of the extra cycle. The
default programming for the extra cycle is set to a CALIBRATION cycle.
EMERGENCY STOP:
This command will immediately stop the analyzer at the current step of the analysis in
progress. The analyzer will go into the MANUAL STOP condition. On the main screen
appear the message "manual stop" red coloured and the FAULT alarm contact will be
activated. To restart the analyzer after an emergency stop, it is necessary to reset the fault by
pressing the yellow colored RESET EM. STOP. This will force the analyzer to stand-by
conditions.