Calibration Menu; Figure 5.9: User Interface At The Calibration Menu - Waltron 3041 User Manual

Silica analyzer
Table of Contents

Advertisement

Cycles ratio: Programs the ratio of the analysis cycles versus the extra cycle (calibration cycle).
For example, 400 analysis to 1 extra cycle means that every 400 analysis cycles the analyzer will
perform 1 calibration cycle.
Cycle wait: Sets the time between analysis cycles when the analyzer is in on-line mode.
Result alarm: Programs the low / high value alarm. The alarm is expressed in % of full scale.
Output Signal: Enables / disables the 4-20 mA output.
Datalog: Enables / disables the internal data logger.
Sample Alarm: Enables / disables the loss of sample alarm.

CALIBRATION MENU

5.8.2
Allows the user to display and set the value of the standard used for calibration and perform
manual blank and calibration cycles (refer to Section 6).

Figure 5.9: User interface at the CALIBRATION Menu.

Waltron User Manual 101-016-B.3
3041 Silica Analyzer
User Interface
38