USER SIGNAL CONNECTIONS
3.6.2
The analyzer provides an External Terminal Block on the outside of the Electronics Enclosure. It
allows the operator to connect to an external device such as a sequencer or sampler, monitor the
4-20mA ouputs and monitor alarm relays. The external terminal block can be configured from
the SERVICE Screen.
Table 3.1 below shows the terminal contact labels starting with pin 1 on the left.
Figure 3.3: Electrical connections for user connections and power connections.
Pin Number
1
2
3
4
5
6
7
8
9
10
11
12
Waltron User Manual 101-016-B.3
3041 Silica Analyzer
Table 3.1: Terminal block pin locations.
Terminal Label
- Input
+ Input
- Signal 2
+ Signal 2
- Signal 1
+ Signal 1
COM ½
OUT Relay 1
OUT Relay 2
OUT Relay 3
OUT Relay 4
COM 3/4
Description
NO contact from an External Device to tell
the analyzer to take a measurement.
4-20mA output for Sample 2 Measurement
4-20mA output for Sample 1 Measurement
Common for Relay 1 and 2
Alarm Relay 1 N/O Contacts
Alarm Relay 2 N/O Contacts
Alarm Relay 3 N/O Contacts
Alarm Relay 4 N/O Contacts
Common for Relay 3 and 4
Installation
28