Installation
WARNING!
Clearance requirements for the analyzer should be about 8" (20 cm) on either side and
about 40" (100 cm) in the front.
Sufficient space for the reagent containers should be provided on the side or beneath the
analyzer.
The reagent containers should be positioned in the reagent bracket or a suitable basin in
case of spills.
NOTE:

MOUTING SCHEMATICS

3.5
10.16" (258 mm)
9.76" (248 mm)

Figure 3.1: The dimensions for mounting the analyzer.

25
The sample drain of the analyzer must be at ambient
pressure with no restriction or counter pressure. Please
verify that this condition has been strictly respected during
installation.
Respect the maximum height of 16" (40 cm) between the
reagent's bottle(s) and the reagent's pump(s).
16.22" (412 mm)
14.96" (380 mm)