4.2. Magnetic sensor (EZU-RV3S)

Magnetic sensor as shown in Fig. 13 is the magnetic sensor for EUP-V53W.
4.2.1. How to attach Magnetic sensor
The procedure of attaching Magnetic sensor is as follow.
1) Confirm that the Magnetic sensor attachment is cleaned, disinfected
and sterilized.
2) Connect the probe, operate the ultrasound diagnostic scanner, and
adjust the image according to the instructions given in the operation
manual for the ultrasound diagnostic scanner.
3) To use RVS (Real-time Virtual Sonography), attach the magnetic sensor
as shown below.
a) Insert Magnetic sensor into the Magnetic sensor attachment with
the correct direction as shown in Fig. 14.
Magnetic sensor attachment
Fig. 14
Never attach the magnetic sensor attachment to the probe in the incorrect
direction, otherwise it may result in false diagnosis.
Direction mark
Fig. 13
Direction mark of
EZU-RV3S
How to attach the Magnetic sensor
Magnetic Sensor
(EZU-RV3S: white)
Magnetic sensor (EZU-RV3S)
Magnetic sensor
(EZU-RV3S)
CAUTION
-12-