Failure
message
Emergency release
Imbalanced load
Check Set Speed
E-01 - E-12
E-13
E-15-E-16
E-17
E-21-E-22
22 |
Troubleshooting
Problem with centrifuge
The centrifuge cannot be operated.
The run does not start or the centrifuge
runs down without being braked.
The centrifuge cannot be operated.
The run does not start or the centrifuge
runs down without being braked.
The centrifuge cannot be operated.
The run does not start or the centrifuge
runs down without being braked.
The centrifuge cannot be operated.
The run does not start or the centrifuge
runs down without being braked.
The centrifuge cannot be operated.
The run does not start or the centrifuge
runs down without being braked.
Temperature sensor broken/controller
defective
Speed for rotor detection exceeded
The centrifuge cannot be operated.
The run does not start or the centrifuge
runs down without being braked.
Possible causes and cures
The lid opens while the device is running.
Close the lid and restart the centrifuge.
If an error message appears again, inform Customer
Service.
Imbalance detected.
Check the load placed in the rotor.
Check that the rotor cross bolts are well greased.
Restart the centrifuge.
If an error message appears again, inform Customer
Service.
The set point speed is higher than the maximum rotor
speed.
Correct the value.
Restart the centrifuge.
If an error message appears again, inform Customer
Service.
Error during the self-test of the centrifuge program and
the electronics.
Restart the centrifuge.
If an error message appears again, inform Customer
Service.
The check sum in the data memory is incorrect.
The software corrects errors automatically.
Check the values of the set point settings, etc.
Restart the centrifuge.
If an error message appears again, inform Customer
Service.
Malfunction in the temperature detection.
Restart the centrifuge.
If an error message appears again, inform Customer
Service.
Restart the centrifuge.
If an error message appears again, inform Customer
Service.
The rotor could not be identified.
Check to see if the rotor is properly installed.
Restart the centrifuge.
If an error message appears again, inform Customer
Service.
Centrifuge GT 2/2R