Service Strategy - Agilent Technologies 54701A User's And Service Manual

2.5-ghz active probe
Table of Contents

Advertisement

Table 6
Recommended Test Equipment
Equipment
Required
Signal Generator
Power Meters (2)
or one Dual-
Channel
Power Sensor (2
Power Splitter
Power Supply
DVM
Power Supply
Adapter/termination N(f)-to-probe, 50 Ω
Adapter
Adapter
Termination
Adapter
C = CalibrationTests, A = Adjustments, T = Troubleshooting
See Also
Critical Specifications
50 MHz to 2.5 GHz
50 MHz to 2.5 GHz,
±3% accuracy
50 MHz to 2.5 GHz, 300 mW
dc to 2.5 GHz, ≤0.2 dB output tracking,
Type-N
Power and control for probe under test 1143A
Resistance ±0.1,% Volts and ohms
±0.01%
5 Vdc
N(f-f), 50 Ω
N(f)-to-BNC(m), 50 Ω
50 Ω, BNC feed-through
BNC (f) to banana (m

Service Strategy

The 54701A Active Probe is a high-frequency instrument with many critical
relationships between parts. For example, the frequency response of the
amplifier on the hybrid is trimmed to match the output coaxial cable. As a
result, to return the probe to optimum performance requires factory repair. If
the probe is under warranty, normal warranty services apply. If the probe is
not under warranty, a failed probe can be exchanged for a reconditioned one
at a nominal cost.
"Troubleshooting and Repair" for further information.
Chapter 3: Service
General Information
Recommended
Use
Model/Part
8663A
C
436A (2),
C
437A (2), or
438A (1)
8482A
C
11667A
C
C,A,T
3458A
C,A,T
6114A
C
11880A
C
1250-1472
C
1250-0077
C,A,T
10100C
C,A,T
1251-2277
C
39