The pulse rate and width may be modified during setup of a Russian or Biphasic Stimulation
treatment or may be modified while a treatment is in progress.
Since a channel pair shares pulse parameters during a Russian or biphasic treatment, you
cannot deliver both a Russian and a biphasic treatments at the same time with this device.
You can deliver two Russian or two biphasic treatments at once.
NOTE: If a biphasic or Russian stimulation treatment is in progress on one channel,
you may not customize the pulse rate and width for the second channel. When you
customize the pulse rate and width, it affects all simultaneous biphasic or Russian stimulation
treatments for both channels (1 and 2).
If you are modifying the pulse rate and width for a treatment in progress, you should also
modify the intensity, as all three of these parameters will affect delivered energy and patient
comfort.
General Information - BIPHASIC & RUSSIAN