• NEVER turn the power on or off while the unit is connected to the patient.
• Always give the Remote Stop device to the patient prior to the treatment.
• Always STOP a treatment before removing or attaching electrodes or leads. Leads and
electrodes must only be applied to the patient before a treatment is started.
• Never use worn or damaged leads or electrodes as these may result in injury to the
patient.
• See "Contraindications, Warnings, and Precautions" in this manual before administering a
treatment to a patient.
High volt electrical stimulation is a pulsed DC current with pulse widths in the microsecond
range and pulse rates ranging from 1 to 200 Hz, with a peak amplitude of up to 1.0 amp. The
Dynatron 850plus and 550plus deliver high volt utilizing a twin-peak monophasic waveform.
High volt treatments with the Dynatron
550plus and 850plus may be delivered using
electrodes or an optional hand-held probe.
The device provides a dedicated channel for
high volt treatment (HV). This channel is
used to deliver high volt treatments only.
The device's other output channels (CH1-2)
remain available for other stim treatments
simultaneously with the high volt treatment.
Set Up High Volt Treatment with
Electrodes
This treatment setup utilizes a double-output
lead with two electrodes; an active and a
dispersive electrode. The size of the dispersive electrode is recommended to be double the
area of the active electrode. If desired, the active output of the lead may be bifurcated to
attach additional active electrodes; however, the combined total area of the active electrodes
should be no more than half the area of the single passive electrode. These setups are
illustrated on this page and the following page.
Dynatron
®

High Volt Instructions

WARNING
850plus & 550plus
Dispersive electrode:
Attach to NEGATIVE
lead output.
DOUBLE
OUTPUT LEAD:
Connect to "HV"
channel
High Volt Electrode Setup
Treatment Instructions - HIGH VOLT
35
( – )
( + )
Active electrode:
Attach to POSITIVE
lead output.