Check to see if conditions may have caused extreme moisture condensation in the device.
This could occur when the machine has become very cold then is brought indoors to a
warm, humid environment. Condensation is a not a serious condition. Allow the
machine to sit in a dry environment until the condensation dries. The machine will
operate normally once the condensation is gone.
ERROR 13: The Error 13 message can occur if you plug in a soundhead while the device is
on. If you get this error, make sure the soundhead is firmly plugged in, then turn the device
off and on again. The message should be cleared.
If you have tried all of these suggestions, and the error is still displayed on the LEDs, the
device may require service by the manufacturer. Make a note of the error message and the
sequence of events that cause the error, and contact Dynatronics Customer Service at 1-800-
874-6251 for further assistance. Do not send the device to Dynatronics without first
contacting the Customer Service Department.