8.

Program parameter settings

The actual software version supports up to four phases usable for program design. During
program run the phases are processed one by one. A program consists of at least one phase.
Phase Parameter
Speed [rpm]
Dwell Time [s]
Final Temp [°C]
dT/dt [K/min]
AFAP
Stop program
General Parameter
Zero NTU as calibrated (xxx)
Auto Calib Zero NTU at start
Automatic CP detection
comments
Program
Load and Save File
Table 1 Program Parameter Settings
Function
The rotational speed of magnetic field that drives
the magnetic bar
Dwell time at target temperature, it starts when
the sample temperature is closer than 0.5 K to
the final temperature
Final temperature or target temperature of that
phase
Slope of the temperature ramp, when not AFAP
The final temperature is achieved as fast as
possible
Defines this phase as the last
Function
Uses pre-calibration of Zero NTU,
recommended for samples which have initial
turbidity at the start of the program,
Refer to
11.5.
Zero calibration
for further details
Recommend setting for samples which are clear
initially. An auto calibration is performed for Zero
NTU
Automatic cloud point detection is for future use
To be printed out on cloud point report
(optionally)
Select one of eight programs to be edited
Load and Save single programs from and to a
file
Turbitronic P
17