Principle of temperature measurement

The Turbitronic P determines the sample temperature in direct contact with sample under test. A
pressure tight platinum temperature probe is placed into the glass tube for best thermal
conductivity.
Figure 6 Glass measuring cell with embedded platinum temperature probe
When the sample cell is inserted into the measuring chamber the temperature value is
transmitted wirelessly to the Turbitronic P.
Figure 7 Infrared wireless connection to the instrument