Replacing The Column Heater - Thermo Scientific Dionex ICS-2000 Operator's Manual

Ion chromatography system
Hide thumbs Also See for Dionex ICS-2000:
Table of Contents

Advertisement

10. Push the cables into the opening in the component panel.
11. Align the top of the suppressor
with the guide line printed on the
component panel (see
Figure
5-20) and align the slots
on the rear of the suppressor
with the tabs on the panel.
12. Press the suppressor onto the
tabs and then slide it down until
it locks into place. Pull slightly
on the center of the suppressor to
verify that it is securely fastened.
Note: Some suppressors require
more force to secure them onto
the tabs.
13. Connect the two eluent and two
regenerant lines to the new
suppressor.
14. Close the ICS-2000 front door.

5.11 Replacing the Column Heater

1. Turn off the ICS-2000 power.
2. Open the front door of the ICS-2000.
3. Follow the steps below to remove the existing column heater:
a. Unscrew the two thumbscrews on the heater cover (they remain attached
to the cover).
b. Pull the heater cover straight out to remove it.
Doc. 031857-03 4/06
3
1
1
Figure 5-20. Suppressor Guidelines and Mounting
Tabs on Component Panel
5 • Service
1
Suppressor
4
Mounting Tabs
2
Suppressor
Cable Connector
3
Guideline for
AES
4
Guideline for
SRS
2
1
5-29