ICS-2000 Ion Chromatography System
5.3

Replacing Tubing and Fittings

The ICS-2000 is plumbed with the tubing and tubing assemblies listed below.
Tubing Size and Type
0.125-mm (0.005-in) ID PEEK
(P/N 044221)
0.25-mm (0.010-in) ID PEEK
(P/N 042690)
0.50-mm (0.020-in) ID PEEK
(P/N 042855)
0.75-mm (0.030-in) ID PEEK
(P/N 044777)
1.58-mm (0.062-in) ID Teflon
(P/N 014157)
25 µL sample loop
(P/N 042857)
10-32 fittings (P/N 043275) and ferrules (P/N 043276) are used for most
tubing connections. For tightening requirements, refer to Installation of
Dionex Liquid Line Fittings (Document No. 031432). The manual is on the
Dionex Reference Library CD-ROM (P/N 053891).
1/8-in fittings (P/N 052267) and ferrules (P/N 048949) are used for
connections to the suppressor
5-14
Color
Used For
Red
Connection from pump pulse damper
to pressure transducer
Black
Connections between other system
components
Orange
Connection from injection port to
injection valve
Green
Connection from injection valve to
waste
Clear
Connection from pump to degas or
eluent reservoir; pump waste
Orange
Connection between ports
on the injection valve
L (4)
port and the eluent reservoir.
REGEN OUT
L (1)
and
Doc. 031857-03 4/06