Overview; Specifications; Performance; Sample Operating Conditions - Waltron 3048 User Manual

Dissolved iron analyzer
Table of Contents

Advertisement

Overview

1 OVERVIEW

SPECIFICATIONS

1.1

PERFORMANCE

1.1.1
Sensor Classification:
Application:
Power Supply:
Humidity:
Ambient Temperature Range:
Range:
Accuracy:
Unit Dimensions:
Positioning and Installation
Details:
Response Time (approx.):
Alarms:
Reproducibility:
Degree of Protection:
Required Maintenance:
Reagent Consumption
(approx.):
Materials in Contact with
Sample:
CE Certification (on request):
Sample Conditioning
Requirements:
Weight:

SAMPLE OPERATING CONDITIONS

1.1.2
Temperature Range:
Inlet Sample:
Outlet Sample:
Turbidity:
Color:
PH:
9
Colorimetric with glass flowcell and LED source
Demineralized, boiler, potable, surface and waste
water
110-220Vac, 50-60 Hz 80 VA
Up to 90% not condensable
10-45º C analyzer (50-113º F)
0 – 100ppb, 0 – 300ppb, 0 – 3ppm*, 0 – 30ppm*
*By Auto Dilution
+/– 2% of full scale
Height=23.6"(60 cm), Width=15"(38 cm),
Depth=8.3"(21 cm)
Wall mounting or with bench support, panel mount
optional
10min, batch process, programmable cycle frequency
2 or 4 configurable alarm relays
+/– 2% of reading, whichever is greater
IP55 (NEMA 4): Wet section, IP65 (NEMA 4x):
Electronics
Monthly replenish reagents, Quarterly replace tubing
1 reagent, 1L per month
Glass, Silicone, Plexiglass, Stainless Steel AISI 316
Meets low voltage and low electromagnetic
compatibility directives
Filtering between 10 and 60 micron, depending on the
matrix, needed only to avoid clogging
17 kg
5-55º C (41-131º F)
100–500 ml/min, atmospheric, flexible tubing I.D. ¼" (6 mm)
Atmospheric, waste tubing O.D. 3/8" (9 mm)
Not applicable, sample blank correction
Not applicable, sample blank correction
3-12
Waltron User Manual 101-024-B.1
3048 Dissolved Iron Analyzer