Siemens Sitrans Probe LU Instruction Manual page 97

Hide thumbs Also See for Sitrans Probe LU:
Table of Contents

Advertisement

Oxley Creek WWTP Water Reclamation
(Probe LU Manual) Vendor Manual
2.7.4.2. Sampling down
2.7.4.3. Window
2.7.4.4. Window Used
2.7.5. Range
2.7.5.1. Near Range (LOW RANGE): (default value 0.25 m)
2.7.5.2. Far Range (HI RANGE): (default 6 m or 12 m, depending on unit)
1.
This parameter is for use only by Siemens Milltronics service technicians.
2.
The transducer face.
7ML19985JB02
Once echo is outside the window, a specified number of consecutive
echoes must appear below the echo locked onto, before measurement is
accepted as valid.
1
(default 0)
A "distance window" centered on the echo, used to derive the reading.
When a new measurement is in the window, the window is re-centered
and the reading is calculated (see
on page 94 for an illustration).
When the value is 0, the window is automatically calculated after each
measurement.
• For slower Measurement Response values, the window is narrow.
• For faster Measurement Response values, the window becomes
progressively wider.
Note:
The echo lock window is stored as standard sample, but
displayed in the engineering units selected to display level. Any value
entered for the echo lock window will be rounded to the nearest
sample.
1
The actual size of the window (whereas
size of the window.
Minimum distance from the reference point
should be considered valid. (This is sometimes referred to as
Hand programmer values
Note:
Far Range can extend beyond the bottom of the tank.
Maximum distance from the reference point
should be considered valid.
Hand programmer values
SITRANS Probe LU (PROFIBUS PA) – INSTRUCTION MANUAL
1
(default 2)
Range: to
Range: to
Doc Id: VM11
Active Date: 15/12/2006
refers to the designed
2
beyond which an echo
m
2
, within which an echo
)
.
Page 91