Siemens Sitrans Probe LU Instruction Manual page 96

Hide thumbs Also See for Sitrans Probe LU:
Table of Contents

Advertisement

Oxley Creek WWTP Water Reclamation
(Probe LU Manual) Vendor Manual
2.7.3.8. Sound Velocity at 20
2.7.3.9. Offset and Velocity calibration
2.7.4. Echo Sampling
2.7.4.1. Sampling up
1.
This parameter is for use only by Siemens Milltronics service technicians.
Page 90
SITRANS Probe LU (PROFIBUS PA) – INSTRUCTION MANUAL
Sound velocity through the medium at 20
344.1 m/s or 1129 ft/s.)
1.
Offset calibration
You can specify a fixed value to apply to all measurements. Use this
function if the reported level is consistently high or low by a fixed
amount (stored in Sensor Offset).
Open menu
option to change calibration, select the calibration type, then enter
new calibration value.
2.
Sound Velocity calibration
Note:
If a velocity calibration is performed after the sensor offset
value has been set manually, the sensor offset value will be reset
to 0.
Used to calculate sound velocity when either the acoustic beam
environment is other than air or unknown, or the reading accuracy is
acceptable only at higher material level.
For best results, calibrate with the level at a known value close to
empty, and manually determine the actual distance from sensor to
material level.
Open menu
option to change calibration, select the calibration type, then enter
new calibration value.
Note:
The Echo Sampling parameters are for use by authorized Siemens
Milltronics Service personnel or technicians familiar with Siemens Milltronics
echo processing techniques.
1
Once echo is outside the window, a specified number of consecutive
echoes must appear above the echo locked onto, before measurement is
accepted as valid.
o
C (default 344.13)
(default 5)
Doc Id: VM11
Active Date: 15/12/2006
o
C. (Sound velocity in air is
, select the
, select the
7ML19985JB02

Advertisement

Table of Contents
loading

Table of Contents