6 Be sure to cover the exposed end of the fiber-optic cable.
7 Use a T-10 TORX driver to remove the four screws securing the A2 assembly
to the instrument's right-side frame.
8 Place the instrument on it's right side.
9 Use a T-10 TORX driver to remove the three screws securing the A2 assembly
to the bottom of the instrument's main deck sheet metal assembly.
10 Remove the A2 assembly.
Replacing Instrument Assemblies
Servicing
5-29