c. Select the devices to remove and push OK. Confirm to remove the selected devices from the
telegram list.
Note: Use the drag and drop function to change the elements on the Telegram list. Push and hold on the element
and move the device or tag to the new position. Make sure to update the PLC configuration accordingly after the
elements are moved.

6.5.4 Configure the Modbus TCP

Use the Modbus TCP/IP to connect the controller to measurement and control systems that use the
TCP/IP protocol for data transmission.
Make sure that all of the necessary electrical connections are complete. Make sure that the Ethernet
port in the controller is configured to Modbus TCP. Refer to
Note: To upgrade a controller to have Modbus TCP/IP, refer to
Configure the controller to use the Modbus TCP communication protocol as follows:
1. Push the main menu icon, then select Outputs > Modbus TCP.
2. Set the Modbus TCP button to on.
3. Select an option.
Option
Modbus TCP
IP address
TCP Port
Telegram
Modbus address
Virtual modbus
slave
Data order
Simulation
Description
Sets the Modbus TCP to on or off.
Shows the IP address.
Shows the TCP Port.
Manages the Telegram data structure. Select the devices and the device data tags in
the Telegram. Refer to
Configure the telegram
Sets the address (1 to 255) of the Modbus secondary device configured in the
Telegram menu.
Adds virtual secondary devices. The virtual secondary devices are copies of devices
configured in the Telegram menu. When set to on, the virtual device is enabled. When
set to off (default), the virtual device is disabled.
Sets the byte sequence for the transmission of floating point values. A floating point
value has 4 bytes. The floating point value is only data of the configured slaves.
• Normal (default)—The pairs are not changed.
• Swapped—Changes the first byte pair with the last byte pair.
Note: An incorrect setting in the Data order menu can cause small changes in the
floating point values (moved in one register).
Simulates two floating point values and errors (or statuses) to replace an instrument.
The first floating point value goes through a ramp between limits set in the Minimum
and Maximum menus.
• Simulation—Enables or disables the simulation. Options: on or off.
• Period—Sets the time in minutes (0 to 1000) for the first floating point value to go
through the full range between Minimum and Maximum settings.
• Maximum—The upper limit for the first floating point value. Range: 0.0 to 1000.0
• Minimum—The lower limit for the first floating point value. Range: 0.0 to 1000.0
• Error—Sets the value in the first simulated register. Range: 0 to 65535
• Status—Sets the value in the second simulated register, Range: 0 to 65535
• Toggle—Changes the direction of the simulated ramp application.
LAN connection
on page 25.
Enter license key
on page 43.
on page 38.