8 Diagnosis, asset management and service
Failure
Code
Text message
F013
No valid measured value
available
F017
Adjustment span too small
F025
Error in the linearization table
F036
no operable sensor software
F040
Error in the electronics
F041
Communication error
F080
General software error
F105
Measured value is deter-
mined
F113
Communication error
F260
Error in the calibration
F261
Error in the instrument set-
tings
F264
Installation/Setup error
F265
Measurement function dis-
turbed
44
maintenance for the instrument because a failure is expected in the
near future (e.g. due to buildup).
This status message is inactive by default.
Cause
Gauge pressure or low pressure
Measuring cell defective
Adjustment not within specification
Index markers are not continuously ris-
ing, for example illogical value pairs
Failed or interrupted software update
Hardware defect
No connection to the sensor electronics Check connection between sensor and
General software error
The instrument is still in the switch-on
phase, the measured value could not
yet be determined
Error in the internal instrument commu-
nication
Error in the calibration carried out in the
factory
Error in the EEPROM
Error during setup
Error when carrying out a reset
Inconsistent settings (e.g.: distance, ad-
justment units with application process
pressure) for selected application
Invalid sensor configuration (e.g.: ap-
plication electronic differential pressure
with connected differential pressure
measuring cell)
Sensor no longer carries out a meas-
urement
Rectification
Exchange measuring cell
Send instrument for repair
Change the adjustment according to
the limit values
Check linearization table
Delete table/Create new
Repeat software update
Check electronics version
Exchanging the electronics
Send instrument for repair
Exchanging the electronics
Send instrument for repair
main electronics (with separate version)
Disconnect operating voltage briefly
Wait for the end of the switch-on phase
Disconnect operating voltage briefly
Send instrument for repair
Exchanging the electronics
Send instrument for repair
Repeat setup
Repeat reset
Modify settings
Modify connected sensor configuration
or application
Carry out a reset
Disconnect operating voltage briefly