Peak value, pressure
Peak value, temperature
Simulation
Date/Time
VEGABAR 86 • Modbus and Levelmaster protocol
6 Set up the sensor with the display and adjustment module
service purposes. You can find the error codes with description, rea-
son as well as rectification in chapter " Asset Management".
The respective min. and max. measured values are saved in the
sensor. The two values are displayed in menu item " Peak values,
pressure".
In another window you can carry out a reset of the peak values
separately.
The respective min. and max. measured values of the measuring cell
and the electronics temperature are stored in the sensor. In menu
item " Peak value, temperature", both values are displayed.
In another window you can carry out a reset of the two peak values
separately.
In this menu item you simulate measured values. Hence, the signal
path can be tested via the bus system to the input card of the control
system.
Select the requested simulation variable and set the requested value.
To deactivate the simulation, you have to push the [ESC] key and
confirm the message " Deactivate simulation" with the [OK] key.
Caution:
During simulation, the simulated value is output as digital signal.
The status message along with the Asset Management function is "
Maintenance".
Information:
The sensor terminates the simulation automatically after 60 minutes.
6.5.4 Additional adjustments
In this menu item, you adjust the internal clock of the sensor. There is
no adjustment for summer/winter (daylight saving) time.