Quickie
ACCESS
W. GAS SPRING ADJUSTMENT
Refer to the tables below for Gas Spring Adjustment.
Determine correct confi guration based on user's weight. Refer to table below and Fig. 59.
Confi guration
1
up to 150 lbs (up to 68 kg)
2
151-180 lbs (69-82 kg)
3
181-210 lbs (83-95 kg)
4
211-240 lbs (96-109 kg)
5
241-270 lbs (110-122 kg)
6
271-300 lbs (123-136 kg)
Determine correct part based on Tilt Confi guration. Refer to table below and Fig. 60.
Tilt Confi guration
-5° to 20°
0° to 25°
X. INSPECT
Once the wheelchair is assembled and adjusted, it should roll smoothly and easily. All accessories should
also perform smoothly. If you have any problems, follow these procedures:
1.
Review assembly and adjustment sections to ensure the chair was properly prepared.
2.
Review troubleshooting guide.
3.
If your problem persists, contact your Sunrise Medical authorized dealer. If you still have a problem
after contacting your authorized dealer, contact Sunrise Medical customer service. See the
introduction page for details on how to contact your Sunrise Medical authorized dealer or Sunrise
Medical customer service.
252849 Rev. B
59
60
User's Weight
(p/n 251002)
User's Weight
up to 300 lbs (136 kg)
up to 300 lbs (136 kg)
IX. DEALER SERVICE AND ADJUSTMENT
Spring
Position
1
2
1
(p/n 251003)
R
R
R
R
R
R
R
R
Part Number
253448
252535
28
Position 1
2
Position 2
R
R
R
R
p/n 253448:
-5° to 20°
59
60
p/n 252535:
0° to 25°